Support Center > Knowledge base> Article: When do I need to recalibrate the SmartLog Pro® 2 and how can I do it?

When do I need to recalibrate the SmartLog Pro® 2 and how can I do it?

Article ID: 108788 Print
Question
When do I need to recalibrate the SmartLog Pro® 2 and how can I do it?

Answer

Frequency of re-calibration should be based on the critical nature of those ESD sensitive items handled and the risk of failure for the ESD protective equipment and materials. In general, EMIT recommends that calibration be performed annually.

Use the EMIT 50424 Limit Comparator and EMIT 50784 5-Pound Electrode to perform periodic verification (once every 6-12 months) of the SmartLog Pro® 2. The Limit Comparator can be used to check the test limits of the SmartLog Pro® 2 without removing it from the factory. 

See technical bulletin TB-6581 for more information or view a video overview of the process at this link:  https://youtu.be/RTvs6SGOznY

NOTE: Calibration is not required for the temperature and humidity sensor embedded within the SmartLog Pro® 2.

If you wish to send in your tester to our factory for calibration, you will need an approved RMA to include in the box.  The pricing and calibration time for that model are shown here:

https://emit.descoindustries.com/Calibration-Pricing.aspx

We recommend that you fill out and submit the calibration repair form at the link below to request your RMA:

https://emit.descoindustries.com/Calibration-Repair.aspx


related articles

Article Details
Views: 532 Created on: Feb 28, 2023
Date updated: Nov 16, 2023

Poor
Outstanding