Orange arrow What cleaning solution may be used to clean the footplate on a SmartLog?

2466 views
  99% Isopropyl alcohol is the best choice for cleaning because it leaves little to no residue....
Orange arrow How many ESD tests can be stored on the SmartLog Pro® 2?

495 views
  The SmartLog Pro®2 will store thousands of transactions before exceeding its memory. The SmartLog Pro® 2 is intended to be used with the SmartLog Pro Manager Web Software which continuously receives all transactions made at the tester, thus freeing up any memory being used. Internal flash memory al...
Orange arrow Can the SmartLog Pro® 2 be connected to an external monitor?

335 views
  Yes. It has a VGA Port that outputs a mirrored image of the tester’s display to an external monitor. The native video resolution is 800 x 480....
Orange arrow Is the SmartLog Pro Manager Web Software compatible with SmartLogs older than the SmartLog Pro® SE?

364 views
  SmartLog Pro® Manager is only compatible with the SmartLog Pro® SE and SmartLog Pro® 2 models. SmartLog Pro® models may be converted into SmartLog Pro® SE models with the purchase of a new SD card that has the correct programming. Please contact the EMIT Factory for more details....
Orange arrow Does the SmartLog Pro®2 have Wi-Fi capabilities?

554 views
  Yes. Use the Wi-Fi setup to wirelessly connect the SmartLog Pro® 2 to the network in instances where a direct Ethernet cable is not used....
Orange arrow How do you configure the proximity reader on the SmartLog Pro®2?

220 views
  The SmartLog Pro® 2 contains an rf IDEAS WAVE ID®Plus reader capable of reading virtually any proximity or contactless card type in the world. The dual-frequency reader has the ability to hold four user-defined ID card configurations. You may use the procedure in the video linked below to configure...
Orange arrow When do I need to recalibrate the SmartLog Pro® 2 and how can I do it?

519 views
  Frequency of re-calibration should be based on the critical nature of those ESD sensitive items handled and the risk of failure for the ESD protective equipment and materials. In general, EMIT recommends that calibration be performed annually. Use the EMIT 50424 Limit Comparator and EMIT 50784 5...
Orange arrow How do you connect the SmartLog Pro®2 to an Alvarado EDC turnstile?

230 views